Concordance of Near-Infrared Autofluorescence in Parathyroid Glands with Definitive Histopathology Result

Main Article Content

Gustavo Armando Santillan Cisneros
Yuri Jiménez Caprielova
Emmanuel Gonzalez Tello
Grecia Jazmín García Gutiérrez

Abstract

Introduction: Hypoparathyroidism is the most frequent complication of thyroidectomy, which is influenced by multiple factors, mainly unintentional resection, due to its location, size and difficulty in identification, leading to post-surgical hypocalcemia in 20 to 30% of which 1 to 3% of cases, hypocalcemia may be permanent; due to this techniques have been described for its identification,  Benmiloud et al. conducted a case-control study where they compared the efficacy of near-infrared light in detecting parathyroid glands. Parathyroid identification rates were higher in the near-infrared light group compared to the control group (76.3% vs 65.7%).


Material and Methods: Cross-sectional, prospective, comparative and analytical study in a third-level hospital in Mexico City, in 20 patients diagnosed with primary and secondary hyperparathyroidism undergoing partial or total parathyroid surgery, with the aim of determining the concordance of near-infrared autofluorescence in these glands and the resultatwhich  final histopathological, statistical analysis was performed in SPSS 24, with the Kappa and ji-square coefficient,


Results:A total of 20 patients undergoing parathyroid surgery, the most common histopathological finding with 44.2% was parathyroid hyperplasia, with a sensitivity for autofluorescence with the result of histopathology of 82.4% and specificity of 88.9% with PPV of 93.3% and NPV 72.7% and a Kappa coefficient with a value of 0.67 with a p <0.005.


Conclusions: The concordance of near-infrared autofluorescence in parathyroid glands with the result of histopathology is considerable, with a good overall performance for timely detection of parathyroid tissue

Article Details

How to Cite
Cisneros, Gustavo Armando Santillan, Yuri Jiménez Caprielova, Emmanuel Gonzalez Tello, and Grecia Jazmín García Gutiérrez. 2022. “Concordance of Near-Infrared Autofluorescence in Parathyroid Glands With Definitive Histopathology Result”. International Journal of Medical Science and Clinical Research Studies 2 (08):813-18. https://doi.org/10.47191/ijmscrs/v2-i8-17.
Section
Articles

References

I. Bora Kahramangil, E. B. (2017). The use of near-infrared flourescence imaging in endocrine surcgical procedures. Journal of Surgical Oncology , 848-855.

II. .Emin Kose, B. K. (2019). 3 Heterogenesous and low-intensity parathyroid autofluorescence: Patters suggesting hyperfunction at parathyroid exploration. Surgery, 431-437.

III. Giju Thomas, P. M. (2019). Innovative surgical guidance for label-free real-time parathyroid identification. Suergery , 114-123.

IV. R Ladurner, N. A. (2017). Near-infrared autoflourescence imaging to detect parathyroide glands in thyroid surgery . Royal College of Surgeons , 33-36.

V. Di Marco, A. N., & Palazzo, F. F. (2020). Near-infrared autofluorescence in thyroid and parathyroid surgery. Gland surgery, 9(Suppl 2), S136–S146. https://doi.org/10.21037/gs.2020.01.04

VI. Wang, B., Zhu, C. R., Liu, H., Yao, X. M., & Wu, J. (2021). The Ability of Near-Infrared Autofluorescence to Protect Parathyroid Gland Function during Thyroid Surgery: A Meta-Analysis. Frontiers in endocrinology, 12, 714691. https://doi.org/10.3389/fendo.2021.714691

VII. .Berber, E., & Akbulut, S. (2021). Can near-infrared autofluorescence imaging be used for intraoperative confirmation of parathyroid tissue? Journal of surgicaloncology, 124(7),1008–1013. https://doi.org/10.1002/jso.26603

Most read articles by the same author(s)